Showing posts with label reagent kits. Show all posts
Showing posts with label reagent kits. Show all posts

Reagent Strips

By Jeffrey V. de Guzman

Routine chemical examination of the urine has changed dramatically since the early days of urine testing, owing to the development of the reagent strip method for chemical analysis.  Reagent strips currently provide a simple, rapid means for performing 10 medically significant chemical analyses, including pH, protein, glucose, ketones, blood, bilirubin, urobilinogen, nitrite, leukocytes, and specific gravity.

Testing methodology consists of dipping the strip completely, but briefly, into a well-mixed urine specimen; removing excess urine by touching the edge of the strip to the container .  Even though this is a simple procedure, improper technique can result in errors.

In addition to the use of correct testing technique, performance of quality control and care of reagent strips must be observed in order to guarantee quality patient care.  All clinical laboratories must have a quality assurance program in place.  Quality assurance program includes procedures for quality control of specimen collection and handling, reagents and test performance, instrument calibration and maintenance, reporting of results, personnel performance and requirements, safety, and documentation that the program is being followed.

Reagent strips must be protected from deterioration caused by moisture, volatile chemicals, heat, and light.  Strips are packaged in opaque containers with desiccant, and when not in use, these bottles should be stored tightly closed in a cool area.  Bottles should not be opened in the presence of volatile fumes.  All bottles are stamped with an expiration date that represents the functional life expectancy of the chemical pads.  This date must be honoured even if there is no noticeable deterioration of the reagents.

Unexpired strips should be tested for chemical reactivity with controls of known negative and abnormal concentrations.  Distilled water is not recommended for use as a negative control because reagent strip chemical reactions are designed to perform at an ionic concentration consistent with urine.  Additional or confirmatory procedures employing different chemical principles must be available for the substances being tested by reagent strip and should be used when questionable results are obtained or, in some instances, to confirm all positive results.

Observing all these procedures in routine urinalysis produces a win win situation between the laboratory and the patients.  Our business is to provide a quality health care for our patients.  But still, some laboratories do not comply with all these standards.  An example of such violations is the cutting of reagent strips into halves.

To quote one of their reason, “We cut urine strips into halves since cut and whole strips will produce equal results, thus minimizing our costs and consequently the charge to our patients”.

I’m sure you are scratching your head right now.  And if you happen to be the inspector, what would be your answer to this?

Reference: Susan King Strasinger: Urinalysis and body fluids.F.A. Davis Company, 1997

Hemoglobin Determination Methods in the Philippines

This article is intended to address the common violation of clinical laboratories with regards to haemoglobin determination.  Majority of the non-hospital based freestanding clinical laboratories specially primary categories do not even perform the test but a result is being released by multiplying a factor (0.333) to the hematocrit of the patient.  Poor patients, they are paying for the test but without knowing that it is actually not being done by the medical technologist involved.  Such thing is very unprofessional and with all the notices and warnings issued by the regulatory officers, this practice still exist.

The approved standard for quantitative blood  haemoglobin determination in the Philippines is by using the hemiglobincyanide (HiCN) or commonly known as the cyanmethemoglobin methodTo perform this method, blood is mixed with Drabkin's solution, a solution that contains ferricyanide and cyanide. The ferricyanide oxidizes the iron in the hemoglobin, thereby changing hemoglobin to methemoglobin. Methemoglobin then unites with the cyanide to form cyanmethemoglobin.  All forms of hemoglobin likely to occur in circulatory blood, with the exception of sulfhemoglobin, are determined. Cyanmethemoglobin produces a color which is measured in a colorimeter, spectrophotometer, or automated instrument.  The color is suitable for measurement in filter as well as in narrow-band spectrophotometers, because its absorption band at a wavelength of 540 nanometers is broad and relatively flat.  The color relates to the concentration of hemoglobin in the blood.  Standards prepared from either crystalline hemoglobin or washed erythrocytes and stored in a brown glass container and in sterile condition are stable for at least nine months (change 2%).

Since very few primary category clinical laboratory are equipped with hemoglobinometer  or a photometer, the regulatory officers consider manual methods which include the Haden-Hausse and Sahli-Hellige methods. In both methods, blood is mixed with dilute hydrochloric acid. This process hemolyzes the red cells, disrupting the integrity of the red cells' membrane and causing the release of hemoglobin, which, in turn, is converted to a brownish-colored solution of acid hematin. The acid hematin solution is then compared with a color standard.

This methods of determination of haemoglobin as haeminchloride (acid haematin) though not recommended because of their unreliability, are sufficient for routine examination provided they are properly done.

Below is the procedure for the Sahli-Hellige method.

Materials Required:
  • ·         Distilled water
  • ·         Sahli-helligehemoglobinometerkit containing:
o   Small bottle of dilute (approx. 0.1N) hydrochloric acid. Prepare this solution by adding 1 ml of concentrated HCl to 99 ml of distilled water. POUR ACID INTO WATER. Replenish this peri-odically–it must be of proper strength.
o   Graduated tube, with a scale on two sides. On one side is the percentage scale, and on the opposite side is the gram scale. The percentage scale reads from 0 to 170. The gram scale reads from 0 to 24.
  • ·          Pipette, marked at the 20 mm 3 level
  • ·         Stirring rod
  • ·         Color comparator, with a window in the side. On the right and left sides of this opening is the color standard for comparison. The center has an open slot to hold the graduated tube.  
Procedure 
  1. With a medicine dropper, place 5 drops of the 0.lN HCl in the bottom of the graduated tube. Place the tube in the color comparator.
  2. Using well-mixed venous blood or fingertip blood, fill the pipette to the 20 mm mark.
  3. Wipe blood from the outside of the pipette. Transfer blood to the Sahli tube. Note the time.
  4. Aspirate distilled water into the pipette two or three times and transfer these washings to the tube.
  5. Shake until the blood is well mixed and the tube is a uniform color.
  6. Add distilled water, drop by drop, each time mixing the solution with the stirring rod. Keep adding water and mixing until the color of the solution matches the standards on either side. Remove the stirring rod from the tube each time before com-paring. Natural light makes more accurate readings possible.
  7. Five minutes after time noted, read the result from the scale on the tube by noting the graduation mark at the lower edge of the meniscus. Read and report both scales.
Related Posts Plugin for WordPress, Blogger...

Followers

Running A Web Design Business

You can run your own web design business from home, from a cafe or from a beach, all you need is a computer and an internet connection.