It has been said, “work not recorded is work not done”. Therefore, such job must have a corresponding job request / report. Preventive maintenance report indicating among other things such as, the nature of work done, parts used, date of repair / PM, and other important and useful information are usually found.
Equipment Record
Records shall be maintained which include the
following:
Name of equipment
Manufacturer’s name, serial number or other unique identification
Date received and placed in service
Condition when received (new, used)
Records of calibration, date of next calibration
Details of maintenance carried out
Current location, where appropriate
Copy of manufacturer’s operating instructions, where available
History of damage, malfunction or repair
Routine Inspection
Ocular check of the physical surface
Checking electrical plugs, sockets, cables, bulbs
Check screws and other mechanical parts
Check optical elements from mechanical damage and dirt
Check illumination by plugging it to electric power source
Performance Check
Run the field stop diaphragm throughout its full range
Move all the blades of iris by its lever control
Run the condenser height adjustment full range; check its centering screw if it moves the condenser smoothly.
Check the mechanical stage if it can move the specimen smoothly and without play throughout its full range. Check by observing at the microscope’s highest magnification.
Parameter: 1 micron stage movement observed at 1000X magnification looks like 1 mm viewed without a microscope and at a distance of 250 mm.
Objectives mounting shoulder, nosepiece must be absolutely clean. If not, centering and focusing are lost.
Rotate nosepiece through 360º and observe all clicking stop positions from either side without any play
Run the fine and coarse adjustment knobs throughout their full range of adjustment.
Make sure that the upper surface of the stand and the corresponding mounting surfaces of the observation tube must be cleaned & free of corrosion.
Adjustment of interpupillary distance of the tube must run smoothly and without exerting undue force.
All optical surfaces must be absolutely cleaned. Pay special attention to: light bulb, cover glass on field stop, condenser front lens, objective front lens, upper surface of eye lens. Inspect and clean accordingly.
A voice coming out from the four corners of a laboratory, expressing thoughts and ideas, personal experiences, news, and other things inside and outside of the Clinical Laboratory. Also, a collection of events and lessons regarding Malaria Microscopy and Quality Assurance in the Philippines.
Subscribe to:
Post Comments (Atom)
Followers
Running A Web Design Business
You can run your own web design business from home, from a cafe or from a beach, all you need is a computer and an internet connection.
nice blog !! i was looking for blog related of Medical Microscopes manufacturer then i found your blog. it's having valuable and useful information.
ReplyDeleteThe best of both worlds, CILIKA’s benchtop series is a digital microscope with a traditional body. It brings together the benefits of digitization, the comfort of ergonomic design, and the reliability and sturdiness of a traditional setup. With its adjustable variable angle feature, users can work for longer and continuous hours without occupational hazards.
ReplyDelete